Using patented tunable diode laser absorption spectroscopy (TDLAS) technology, the SS2100 gas analyzer measures specific gases (H2O, CO2, H2S, NH3, and C2H2) with precision while avoiding interferences that are common with traditional infrared analyzers. The SS2100 analyzer requires little maintenance and does not need recalibration or periodic replacement parts.
Gas analysis: ppb level measurement of H2O & NH3, ppm level measurement of H2O, H2S, CO2, NH3, and C2H2, and percent (%) level measurement of H2S and CO2
Natural gas: production, storage, transportation, distribution
Check out the findings from the GERG report on the use of a Raman Rxn4 analyzer and Raman Rxn-41 cryogenic probe for the reliable chemical composition measurement in LNG custody transfer applications.
An advanced gas analyzer for hydrogen sulfide (H2S) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels
1
Analyte and Measurement ranges
H2S (Hydrogen Sulfide): 0 to 10 ppmv 0 to 1000 ppmv 0 to 5000 ppmv 0 to 5% volume
Measured Variables
Concentration Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs: 2, Flow Alarm, Validation Request
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Product Headline
An advanced gas analyzer for moisture (H2O) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels
1
Analyte and Measurement ranges
H2O (Moisture): 0 to 10 ppmv 0 to 100 ppmv 0 to 50 ppmv 0 to 5000 ppmv
Measured Variables
Concentration Moisture Dew Point Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs: 2, Flow Alarm, Validation Request
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Product Headline
An advanced gas analyzer for carbon dioxide (CO2) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels
1
Analyte and Measurement ranges
CO2 (Carbon Dioxide): 0 to 10 ppmv 0 to 1000 ppmv 0 ppmv to 5000 ppmv 0 to 5% volume 0 to 10% volume 0 to 20% volume
Measured Variables
Concentration Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs: 2, Flow Alarm, Validation Request
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Product Headline
An advanced gas analyzer for ammonia (NH3) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels
1
Analyte and Measurement ranges
NH3 (Ammonia): 0 to 5 ppmv
Measured Variables
Concentration Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ±10%, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs: 2, Flow Alarm, Validation Request
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Product Headline
An advanced gas analyzer for acetylene (C2H2) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels
1
Analyte and Measurement ranges
C2H2 (Acetylene): 0 to 5 ppmv (minimum) 0 ppmv to 3000 ppmv (maximum)
Measured Variables
Concentration Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs: 2, Flow Alarm, Validation Request
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Product Headline
Advanced multi-channel gas analyzer for hydrogen sulfide (H2S) and moisture (H2O), also known as a 2-Pack. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels
2
Analyte and Measurement ranges
H2S (Hydrogen Sulfide): 0 to 10 ppmv (minimum) 0 to 1000 ppmv (maximum) H2O (Moisture): 0 to 100 ppmv (minimum) 0 to 5000 ppmv (maximum)
Measured Variables
Concentration Moisture Dew Point Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz
Communication
Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (H2O) Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs (H2S): 2, Flow Alarm, Validation Request Digital Outputs (H2O): 2, Concentration Alarm, General Fault
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Product Headline
Advanced multi-channel gas analyzer for hydrogen sulfide (H2S) and carbon dioxide (CO2), also known as a 2-Pack. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels
2
Analyte and Measurement ranges
H2S (Hydrogen Sulfide): 0 to 10 ppmv (minimum) 0 to 1000 ppmv (maximum) CO2 (Carbon Dioxide): 0 to 5000 ppmv 0 to 5% volume 0 to 10% volume 0 to 20% volume
Measured Variables
Concentration Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz
Communication
Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (CO2) Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs (H2S): 2, Flow Alarm, Validation Request Digital Outputs (CO2): 2, Concentration Alarm, General Fault
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Product Headline
Advanced multi-channel gas analyzer for hydrogen sulfide (H2S), moisture (H2O), and carbon dioxide (CO2), also known as a 3-Pack. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels
3
Analyte and Measurement ranges
H2S (Hydrogen Sulfide): 0 to 10 ppmv (minimum) 0 to 1000 ppmv (maximum) H2O (Moisture): 0 to 100 ppmv (minimum) 0 to 5000 ppmv (maximum) CO2 (Carbon Dioxide): 0 to 5000 ppmv 0 to 5% volume 0 to 10% volume 0 to 20% volume
Measured Variables
Concentration Moisture Dew Point Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz
Communication
Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (H2O/CO2) Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs (H2S): 2, Flow Alarm, Validation Request Digital Outputs (H2O/CO2): 2, Concentration Alarm, General Fault
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Real-time quality measurements in biogas and biomethane
PU (Publication)
English version - 10/2023
New version available in English
A discussion on using TDLAS / QF analyzers (SS500, J22, JT33, SS2100, and OXY5500) to measure the concentration of compounds in biogas and biomethane in the Oil & Gas industry.
TDLAS and QF analyzers application guide - Accurate and
reliable measurement of H2O, H2S, CO2, C2H2, NH3, and O2
PU (Publication)
English version - 11/2023
New version available in English
A guide to TDLAS and QF technology application notes (natural gas, hydrogen, biomethane, carbon capture, LNG, gas processing, NGL fractionation, petrochemical, syngas, ammonia, and refining) for the oil & gas and chemical industries (relating to J22, JT33, SS500, SS2100, SS2100a, SS2100i, and OXY5500 analyzers).
Factory acceptance test (FAT) information - For
TDLAS and QF analyzers
PU (Publication)
English version - 10/2023
New version available in English
An overview about the factory acceptance tests (FATs) offered for TDLAS and QF analyzers (SS2100, SS2100i, SS2100a, J22, JT33, SS500, OXY5500) to customers.
An overview about the engineered approval drawings offered for TDLAS and QF analyzers (SS2100, SS2100i, SS2100a, J22, JT33, OXY5500, SS500) to enable customers to confirm that the system fabrication meets their approval.